No comment |
>
Instrumentation >
Digital multimeter >
Multimeter 4 1/2 Logging of True-rms digital VC189 VICTORPLCAGE:Multimeter 4 1/2 Logging of True-rms digital VC189 VICTOR
Characteristics
l Measurements
AC voltage, DC voltage, AC current ,DC current, OHM, Frequency, Period, Pulse width,Capacitance, Beep, Diode, dBm, Thermocouple (TC), Thermal resistance (RTD), AC+ DC, AC + Hz measurement, Maximum, Minimum, Average measurement (MAX / MIN / AVG), The relative value measurement (REL_%)
l Measurements show that 55,000 words, the basic accuracy 0.02% (VICTOR 189) / 0.05% (VICTOR 187)
l Three measurement rate: FAST, SLOW, SMOOTH, the fastest refresh rate of simulation bar: 14 / sec
l Manual range or automatic range selection
l Measuring data show that hold functional (DIS_HOLD) and automatically hold functions(AUTO_HOLD)
l Peak measurements, 500us peak sampling capture, capable of capturing crest Pulse
l AC voltage and AC current are true rms measurement, AC voltage bandwidth 20Hz~30KHz(VICTOR 187),20Hz-100KHz(VICTOR 189), AC current bandwidth 20Hz-30KHz
l Option 1 ~ 2400 Ω reference impedance measurement of decibels
l Low pass filter function normally by the RP-raising and the frequency of the electrical waveform measurement results
l Can be configured clamp to measure the large current
l The eight kinds of thermocouple measurement, the two kinds of heat resistance measurement, high-precision cold-automatic compensation, ? or ? temperature display
l Large-screen display and more data can be displayed at the same time that is measurements and other information related to the measurement
l Built-in real-time clock, for the record and provide accurate measurement of time
l Easy to operate multi-parameter data record: manual records and records of the incident, and easy to query the data records
l Adopted panel calibration technology, can be done without opening the case calibration
l Use alkaline batteries, Ni-Hi battery-powered, the battery and fuse can be easily replaced.
l Backlight is automatically closed and automatic power off function
l Large-screen LCD display, with white LED backlighting
l The meter can communicate with computer by isolation interface USB
l Through the friendly interface, users can easily access the data of instrument in a number of parameters, and may be the data storage, processing, management, access to the data tables and other graphics or display
l Simple man-machine operating, delicate, strongly and suitable for field use
Specifications for products
Measurements | Range | Measuring range | Resolution | ||
DCV | 5.0000V | 0.0001V | Input impedance: 10MΩ | ||
50.000V | 0.001V | ||||
500.00V | 0.01V | ||||
1000.0V | 0.1V | ||||
DCmV | 50.000mV | 0.001mV | Input impedance: approx. 100MΩ | ||
500.00mV | 0.01mV | ||||
2200.0mV | 0.1mV | ||||
ACV | 5.0000V | 0.0001V | Input impedance: 10MΩ / 50pF | CF<3 187: 20Hz~30kHz (10~110% Range) 189: 20Hz~100kHz (5~110% Range) | |
50.000V | 0.001V | ||||
500.00V | 0.01V | ||||
760.0V | 0.1V | ||||
ACmV | 50.000mV | 0.001mV | Input impedance: 1MΩ / 50pF | ||
500.00mV | 0.01mV | ||||
OHM | 500.00Ω | 0.01Ω | Open circuit voltage <2.5V | ||
5.0000KΩ | 0.0001KΩ | ||||
50.000KΩ | 0.001KΩ | ||||
500.00KΩ | 0.01KΩ | ||||
5.0000MΩ | 0.0001 MΩ | ||||
50.000MΩ | 0.001 MΩ | ||||
500.0MΩ | 0.1 MΩ | ||||
DCA | 5.0000A | 0.0001 A | Voltage Drop <0.04 V/A | ||
20.000A | 0.001 A | ||||
DCmA | 50.000mA | 0.001mA | Voltage Drop <1.8 mV/mA | ||
500.00mA | 0.01 mA | ||||
DCuA | 500.00uA | 0.01 uA | Voltage Drop 103uV/uA | ||
5000.0uA | 0.1 uA | ||||
5.0000A | 0.0001 A | ||||
ACmA | 50.000mA | 0.001mA | |||
500.00mA | 0.01 mA | ||||
ACuA | 500.00uA | 0.01uA | |||
50000uA | 0.1uA | ||||
FREQ | 9.9999Hz | 0.0001Hz | |||
99.999Hz | 0.001Hz | ||||
999.99Hz | 0.01Hz | ||||
9.9999KHz | 0.0001KHz | ||||
99.999KHz | 0.001KHz | ||||
999.99KHz | |||||
0.01KHz | |||||
9.9999MHz | 0.0001MHz | ||||
99.999MHz | 0.001MHz | ||||
DUTY | 0.1%~99.9% | 0.1% | 0.5Hz~1KHz, Square Wave | ||
PULSE | 199.99ms | 0.01ms | |||
1999.9ms | 0.1ms | ||||
DIODE | 2.2000V | 0.0001V | Open circuit voltage:<3.5V Short circuit:0.8mA | ||
CHCNT | 500.0Ω | 0.1Ω | Open circuit voltage:<3.5V Short circuit:<0.8mA Short alarm:approx. <20Ω Open alarm:approx. >120Ω | ||
CAP | 9.999nF | 0.001nF | |||
99.99nF | 0.01nF | ||||
999.9nF | 0.1nF | ||||
9.999μF | 0.001uF | ||||
99.99μF | 0.01uF | ||||
999.9μF | 0.1uF | ||||
9.999mF | 0.001mF | ||||
99.99mF | 0.01mF | ||||
TC | R | -40°C~1760°C | 1°C | ITS-90 | |
S | -40°C~1760°C | ||||
K | -270.0~1370.0 | 0.1°C | |||
E | -270.0~720.0 | 0.1°C | |||
J | -210.0~850.0 | 0.1°C | |||
T | -270.0~400.0 | 0.1°C | |||
N | -270.0~1300.0 | 0.1°C | |||
B | 400~1800 | 1°C | |||
RTD | Pt100 | -200.0~850.0 | 0.1°C | Pt100-385 | |
Pt1000 | -200.0~630.0 | 0.1°C |
Other Features:
l DC Voltage measurement
CMRR:50Hz / 60Hz ≥ 120dB
SMRR:50Hz / 60Hz(Rs=1kΩ) ≥ 80dB
l AC Voltage measurement
CMRR:50Hz / 60Hz(Rs=1kΩ) ≥ 80 dB
A residual reading with input shorted, will not affect stated accuracy above 5 % of range.
l OHM measurement
For relative humidity greater than 70 %, resistance accuracy is 0.5 % over 1 MΩand 2.5 % over 10 MΩ.
l Internal temperature compensation sensor RJC:
temperature range: -10 ~ 50?, compensation error ≤ ± 0.5 ?, compensation time for 10 seconds each time
l Maximum voltage between any Terminal and Earth Ground :
1000V DC / AC rms
Fuse protection for mA input: 500 mA / 250V FAST fuse
fuse protection for 20A input: 20 A / 250V FAST fuse
Can be as high as 20 A continuous measurement of the current, if the signal detected in 10 A ~ 20 A of maintaining maximum of 30 seconds, then specified the need to increase the accuracy of 0.5%
l Real-Time Clock
≤ ± 1min / 1month
Temp. Influence 50 ppm per C
l Measured rate
Functional range | Measuring rate | ||
FAST | SLOW | SMOOTH | |
DC voltage and current | 14 | 7 | 1 |
AC voltage and current | 14 | / | 1.8 |
Resistors, diodes, on-off | 7 | / | 1.8 |
General Specifications:
l Operating Temperature and Humidity
0 to 50oC,(no condensation ≤10oC) ≤ 85%RH
l Storage Temperature and Relative Humidity
-20oC to 60oC,≤90%RH
l Altitude
2000m max. above sea level
l Rated transient over-voltage
6KV(1.2*50us)CAT?, CAI ?, CAT ? peak measurement
l Vibration and Shock
Random 2g, 5 to 500Hz
1 meter drop test
l Period of calibration and warm-up time
Accuracy is specified for a period of one year after calibration or other
Warm-up time is 10 minutes or other
l Display and LED
* LCD:36.3×68.0mm
* white LED,able to set time 0~9999S
l Supply
* 4 x 1.5V Alkaline batteries supply
* 4 x 1.2V Ni-Hi batteries supply
* Auto Power-off:able to set time 0~9999minutes
l Safety Standards
IEC 61010-1, 2000
Overvoltage category III (Max. input voltage:AC/DC600V)
Overvoltage category II (Max. input voltage:AC/DC1000V)
Pollution degree 2
l Weight and Size
205 × 95 × 42(mm) ,approximate 500g
PLCAGE:Multimeter 4 1/2 Logging of True-rms digital VC189 VICTOR
Characteristics
l Measurements
AC voltage, DC voltage, AC current ,DC current, OHM, Frequency, Period, Pulse width,Capacitance, Beep, Diode, dBm, Thermocouple (TC), Thermal resistance (RTD), AC+ DC, AC + Hz measurement, Maximum, Minimum, Average measurement (MAX / MIN / AVG), The relative value measurement (REL_%)
l Measurements show that 55,000 words, the basic accuracy 0.02% (VICTOR 189) / 0.05% (VICTOR 187)
l Three measurement rate: FAST, SLOW, SMOOTH, the fastest refresh rate of simulation bar: 14 / sec
l Manual range or automatic range selection
l Measuring data show that hold functional (DIS_HOLD) and automatically hold functions(AUTO_HOLD)
l Peak measurements, 500us peak sampling capture, capable of capturing crest Pulse
l AC voltage and AC current are true rms measurement, AC voltage bandwidth 20Hz~30KHz(VICTOR 187),20Hz-100KHz(VICTOR 189), AC current bandwidth 20Hz-30KHz
l Option 1 ~ 2400 Ω reference impedance measurement of decibels
l Low pass filter function normally by the RP-raising and the frequency of the electrical waveform measurement results
l Can be configured clamp to measure the large current
l The eight kinds of thermocouple measurement, the two kinds of heat resistance measurement, high-precision cold-automatic compensation, ? or ? temperature display
l Large-screen display and more data can be displayed at the same time that is measurements and other information related to the measurement
l Built-in real-time clock, for the record and provide accurate measurement of time
l Easy to operate multi-parameter data record: manual records and records of the incident, and easy to query the data records
l Adopted panel calibration technology, can be done without opening the case calibration
l Use alkaline batteries, Ni-Hi battery-powered, the battery and fuse can be easily replaced.
l Backlight is automatically closed and automatic power off function
l Large-screen LCD display, with white LED backlighting
l The meter can communicate with computer by isolation interface USB
l Through the friendly interface, users can easily access the data of instrument in a number of parameters, and may be the data storage, processing, management, access to the data tables and other graphics or display
l Simple man-machine operating, delicate, strongly and suitable for field use
Specifications for products
Measurements
Range
Measuring range
Resolution
DCV
5.0000V
0.0001V
Input impedance:
10MΩ
50.000V
0.001V
500.00V
0.01V
1000.0V
0.1V
DCmV
50.000mV
0.001mV
Input impedance:
approx. 100MΩ
500.00mV
0.01mV
2200.0mV
0.1mV
ACV
5.0000V
0.0001V
Input impedance:
10MΩ / 50pF
CF<3
187:
20Hz~30kHz
(10~110% Range)
189:
20Hz~100kHz
(5~110% Range)
50.000V
0.001V
500.00V
0.01V
760.0V
0.1V
ACmV
50.000mV
0.001mV
Input impedance:
1MΩ / 50pF
500.00mV
0.01mV
OHM
500.00Ω
0.01Ω
Open circuit voltage
<2.5V
5.0000KΩ
0.0001KΩ
50.000KΩ
0.001KΩ
500.00KΩ
0.01KΩ
5.0000MΩ
0.0001 MΩ
50.000MΩ
0.001 MΩ
500.0MΩ
0.1 MΩ
DCA
5.0000A
0.0001 A
Voltage Drop <0.04 V/A
20.000A
0.001 A
DCmA
50.000mA
0.001mA
Voltage Drop
<1.8 mV/mA
500.00mA
0.01 mA
DCuA
500.00uA
0.01 uA
Voltage Drop
103uV/uA
5000.0uA
0.1 uA
5.0000A
0.0001 A
ACmA
50.000mA
0.001mA
500.00mA
0.01 mA
ACuA
500.00uA
0.01uA
50000uA
0.1uA
FREQ
9.9999Hz
0.0001Hz
99.999Hz
0.001Hz
999.99Hz
0.01Hz
9.9999KHz
0.0001KHz
99.999KHz
0.001KHz
999.99KHz
0.01KHz
9.9999MHz
0.0001MHz
99.999MHz
0.001MHz
DUTY
0.1%~99.9%
0.1%
0.5Hz~1KHz,
Square Wave
PULSE
199.99ms
0.01ms
1999.9ms
0.1ms
DIODE
2.2000V
0.0001V
Open circuit voltage:<3.5V
Short circuit:0.8mA
CHCNT
500.0Ω
0.1Ω
Open circuit voltage:<3.5V
Short circuit:<0.8mA
Short alarm:approx. <20Ω
Open alarm:approx. >120Ω
CAP
9.999nF
0.001nF
99.99nF
0.01nF
999.9nF
0.1nF
9.999μF
0.001uF
99.99μF
0.01uF
999.9μF
0.1uF
9.999mF
0.001mF
99.99mF
0.01mF
TC
R
-40°C~1760°C
1°C
ITS-90
S
-40°C~1760°C
K
-270.0~1370.0
0.1°C
E
-270.0~720.0
0.1°C
J
-210.0~850.0
0.1°C
T
-270.0~400.0
0.1°C
N
-270.0~1300.0
0.1°C
B
400~1800
1°C
RTD
Pt100
-200.0~850.0
0.1°C
Pt100-385
Pt1000
-200.0~630.0
0.1°C
Other Features:
l DC Voltage measurement
CMRR:50Hz / 60Hz ≥ 120dB
SMRR:50Hz / 60Hz(Rs=1kΩ) ≥ 80dB
l AC Voltage measurement
CMRR:50Hz / 60Hz(Rs=1kΩ) ≥ 80 dB
A residual reading with input shorted, will not affect stated accuracy above 5 % of range.
l OHM measurement
For relative humidity greater than 70 %, resistance accuracy is 0.5 % over 1 MΩand 2.5 % over 10 MΩ.
l Internal temperature compensation sensor RJC:
temperature range: -10 ~ 50?, compensation error ≤ ± 0.5 ?, compensation time for 10 seconds each time
l Maximum voltage between any Terminal and Earth Ground :
1000V DC / AC rms
Fuse protection for mA input: 500 mA / 250V FAST fuse
fuse protection for 20A input: 20 A / 250V FAST fuse
Can be as high as 20 A continuous measurement of the current, if the signal detected in 10 A ~ 20 A of maintaining maximum of 30 seconds, then specified the need to increase the accuracy of 0.5%l Real-Time Clock
≤ ± 1min / 1month
Temp. Influence 50 ppm per Cl Measured rate
Functional range
Measuring rate
FAST
SLOW
SMOOTH
DC voltage and current
14
7
1
AC voltage and current
14
/
1.8
Resistors, diodes, on-off
7
/
1.8
General Specifications:
l Operating Temperature and Humidity
0 to 50oC,(no condensation ≤10oC) ≤ 85%RH
l Storage Temperature and Relative Humidity
-20oC to 60oC,≤90%RH
l Altitude
2000m max. above sea level
l Rated transient over-voltage
6KV(1.2*50us)CAT?, CAI ?, CAT ? peak measurement
l Vibration and Shock
Random 2g, 5 to 500Hz
1 meter drop test
l Period of calibration and warm-up time
Accuracy is specified for a period of one year after calibration or other
Warm-up time is 10 minutes or other
l Display and LED
* LCD:36.3×68.0mm
* white LED,able to set time 0~9999S
l Supply
* 4 x 1.5V Alkaline batteries supply
* 4 x 1.2V Ni-Hi batteries supply
* Auto Power-off:able to set time 0~9999minutes
l Safety Standards
IEC 61010-1, 2000
Overvoltage category III (Max. input voltage:AC/DC600V)
Overvoltage category II (Max. input voltage:AC/DC1000V)
Pollution degree 2
l Weight and Size
205 × 95 × 42(mm) ,approximate 500g
No comment |