No comment |
>
Instrumentation >
Digital multimeter >
VICTOR 25 Multi-function Process CalibratorVICTOR 25 Multi-function Process CalibratorDetailed Product Description
Measuring function
Range
Measuring range
Resolving power
Accuracy
Remark
DC Voltage
50mV
-5.000mV55.000mV
1μV
0.02%+10
input impedance: about100MΩ
500mV
-10.00mV550.00mV
10μV
0.02%+2
5V
-0.1000V5.5000V
0.1mV
0.02%+5
input impedance: 1MΩ
50V
0.00V55.00V
10mV
0.03%+5
DC Current
50mA
-5.000mA55.000mA
1μA
0.02%+2
input impedance: 10Ω
OHM
500Ω About 1mA exciting
0.00Ω550.00Ω
0.01Ω
0.02%+0.1Ω
Plough voltage:about 2.5V not including the accuracy of down-lead resistance
5KΩ About 0.1mA exciting
0.0000 KΩ5.5000KΩ
0.1Ω
0.02%+0.5Ω
Frequency
100Hz
2Hz110.0Hz
0.1Hz
±2
input impedance: 100 kΩ or more Square wave at least 1Vp-p 50% duty cycle
1KHz
0.100KHz1.100KHz
1Hz
10KHz
0.10KHz11.0KHz
0. 1KHz
Thermocouple
R
-40°C1760°C
1°C
-4500 : 1.8 501760 : 1.5
Adopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error
S
-20°C1760°C
K
-200°C950°C
0.1°C
-2000 : 1.2 01370 : 0.8
E
-200°C500°C
-2000 : 0.9 01000 : 1.5
J
-200°C700°C
-2000 : 1.0 01200 : 0.7
T
-200°C400°C
-2000 : 1.0 0400 : 0.7
N
-200°C1000°C
-2000 : 1.5 01300 : 0.9
B
400°C1800°C
1°C
400800 : 2.2 8001000 : 1.8 10001800 : 1.4
Thermo resistance
PT100
-200°C700°C
0.1°C
-2000 : 0.5 0400 : 0.7 400850 : 0.8
PT100-385 thermometric scale not including the accuracy of down-lead resistance
PT1000
-200°C630°C
-200100 : 0.22 100300 : 0.3 300630 : 0.4
Cu50
-50°C150°C
-500 : 0.5 0150 : 0.7
Switch mete measure
About 1mA exciting
CLOSE / OPEN
Continuity test
500Ω About 1mA exciting
≤50Ω vocalism
0.01Ω
No comment |