No comment |
VICTOR 03 Thermo Resistance Calibrator
Input
Range
Input range
Resolution
Accuracy
Explanation
Resistance
400Ω
0.0~450.0Ω
0.1Ω
±0.05% Setting value ±0.2Ω
Measurong current 1mA Note1
Thermoresis-tance
Pt100
-200.0~850.0?
0.1?
±0.05% Setting value ±0.6?
1mA exciting current use Pt100-385
Measuring current 1mA Note1.2
Cu50
-50.0~150.0?
0.1?
1mA excitong current Note1.2
Input
Range
Input range
Resolution
Accuracy
Explanation
Resistance
400Ω
0.0~400.0Ω
0.1Ω
±0.05% Setting value ±0.2Ω
±1mA exciting current Note1.2
Thermoresis-tance
Pt100
-200.0~850.0?
0.1?
±0.05% Setting value ±0.6?
±1mA exciting current use Pt100-85
Note1.2
Cu50
-50.0~150.0?
0.1?
±1mA excitong current Note1.2
Power
9V battery (ANSI/NEDA 1640A or IEC 6LR619 alkaline) or AC power adapter
Battery
Approx. 20 hours under the condition of 10mA
Max.allowable voltage
30V
Operation temperture range
0-50°C
Operation humidity range
≤80%RH
Store temperature range
≤-10°C-50°C
Store humidity range
≤90%RH
Size
200×100×40mm
Weight
550g(with holster)
Accessories
Operation manual,test line CF-36(probe with alligator clip)
Options
AC adapter(VCPS),test line CF-31-A(probe clip)
Security
Conforms to IEC 1010
Note:
1.Not including the accessorial lead resistance
2.The range of exciting current:0.5mA~2mA,the max.output voltage≤2V
3.Temperature coefficient?±0.005%?range/°C(0°C-18°C,28°C-50°C)
No comment |