Calibrator VICTOR 15+ VICTORFeatures:
High Accuracy of 0.02% for measure and source. 5 digits display for measure, and 6 digits display for
source.A dual, Liquid Crystal;Display White LED for backlight
TECHNICAL DATA (Output Function)
Function
Reference
Range
Resolution
Accuracy
Remark
DC voltage
1000mV
-100.00~1100.00mV
10μV
0.02+0.01
Maximum output: 2mA
10V
-1.0000~11.0000V
0.1mV
0.02+0.01
Maximum output: 5mA
DC current
20mA
0.000~22.000mA
1μA
0.02+0.02
Simulator transmitter: 5–28V power
supply outside;1KΩ at 20mAFREQ
100Hz
1.00Hz~110.00Hz
0.01Hz
±2 count
Output voltage: +1–+11 V (zero
base waveform);
Amplitude accuracy: ±10%+0.5V;
Maximum load: >100 KΩ;
Duty Cycle: 50%.1KHz
0.100KHz~1.100KHz
1Hz
10KHz
1.0KHz~11.0KHz
0.1KHz
PULSE
100Hz
1~100000cycles
1cyc
——
1KHz
10KHz
LOOP
24V
±10%
Maximum output current: 25 mA
Short circuit protectedMeasuring Function
Function
Reference
Range
Resolution
Accuracy
Remark
DCV
-20.00~220.00mV
10μV
0.02+0.02
-20.00~220.00mV
Input Resistance:approximately 100MΩ
Input Resistance:approximately 1MΩ-0.5000~5.5000V
0.1mV
0.02+0.01
-0.5000~5.5000V
-5.000~55.000V
1mV
0.03+0.01
-5.000~55.000V
DCmA
50mA
-4.000~55.000mA
1μA
0.02+0.01
Input Resistance:5Ω
FREQ
500Hz
3Hz~500.00Hz
0.01Hz
±2count
Input Impedance:100 kΩ at least;
Sensitivity:3Vp-p minimum;
Duty Cycle: 50%.5KHz
3Hz~5.0000KHz
0.1Hz
50KHz
3Hz~50.000KHz
1Hz
CONTINUITY
≤250Ω sound
CLOSE / OPEN
Open circuit voltage:about 2.5V
PRESS
Defined by
pressure module——
Defined by
pressure moduleDefined by pressu-
re moduleFor more detail, refer the pressure
module about APM.
General Characters
Power
4×1.5AAA-size alkaline batteries (AC adapter can charge Ni-Hi battery)
Max.allowable voltage
30V
Operation temperture range
0-50°C
Operation humidity range
≤80%RH
Store temperature range
≤-10°C-55°C
Store humidity range
≤90%RH
Size
205×95×42mm
Weight
500g(with holster)
Accessories
Test leads,Fuse
Options
AC adapter(VCPS),
Security
Conforms to IEC 1010