Temperature Calibrator 5 1/2 VC01 VICTOROutput | Range | Output range | Resolution | Accuracy | Explanation |
Voltage | 100mV | -10.00 ~110.00mV | 0.01mV | ±0.05% Setting value ±30uV | The max. output current ±5mA |
1000mV | -100.00~1100.0mV | 0.1mV | ±0.05% Setting value ± 0.3mV | ||
Resistance | 400Ω | 0.0~400.0Ω | 0.1Ω | ±0.05% Setting value ±0.2Ω | 1mA exciting current (note1.2) |
Thermocouple | R | -40~1760ºC | 1ºC | ±0.05% Setting value ±3ºC (≤100ºC) | Adopt the ITS-90 |
S | -20~1760ºC | 1ºC | ±0.05% Setting value ±2ºC(>100ºC) | temperature standard | |
B | 400~1800ºC | 1ºC | ±0.05% Setting value ±3ºC (400~600ºC) | ||
±0.05% Setting value ±2ºC (>600ºC) | |||||
E | -200.0~1000.0ºC | 0.1ºC | ±0.05% Setting value ±2ºC (≤-100ºC) | (Note 3) | |
K | -200.0~1370ºC | 0.1ºC | |||
J | -200.0~1200.0ºC | 0.1ºC | ±0.05% Setting value ±1ºC(>-100ºC) | ||
T | -200.0~400.0ºC | 0.1ºC | |||
N | -200.0~1300.0ºC | 0.1ºC | |||
Thermoresistance | Pt100 | -200.0~850.0ºC | 0.1ºC | ±0.05% Setting value ±0.6ºC | Adopt the PT100-385 1mA (Note 1.2) |
Cu50 | -50.0~150.0ºC |
Note:
1. Not including the accessorial lead resistance.
2. The range of exciting current:0.5mA~2mA,the max.output voltage≤2V.
3. Not including the accuracy of inner temperature compensatory transducer.
The range of inner temperature compensatory transducer:-10-50°C,compensatory error ≤0.5°C.
4. Temperature coefficient: ±0.005%, range /°C(0°C-18°C,28°C-50°C).
Temperature Calibrator 5 1/2 VC01 VICTOR
Output
Range
Output range
Resolution
Accuracy
Explanation
Voltage
100mV
-10.00 ~110.00mV
0.01mV
±0.05% Setting value ±30uV
The max. output current ±5mA
1000mV
-100.00~1100.0mV
0.1mV
±0.05% Setting value ± 0.3mV
Resistance
400Ω
0.0~400.0Ω
0.1Ω
±0.05% Setting value ±0.2Ω
1mA exciting current (note1.2)
Thermocouple
R
-40~1760ºC
1ºC
±0.05% Setting value ±3ºC (≤100ºC)
Adopt the ITS-90
S
-20~1760ºC
1ºC
±0.05% Setting value ±2ºC(>100ºC)
temperature standard
B
400~1800ºC
1ºC
±0.05% Setting value ±3ºC (400~600ºC)
±0.05% Setting value ±2ºC (>600ºC)
E
-200.0~1000.0ºC
0.1ºC
±0.05% Setting value ±2ºC (≤-100ºC)
(Note 3)
K
-200.0~1370ºC
0.1ºC
J
-200.0~1200.0ºC
0.1ºC
±0.05% Setting value ±1ºC(>-100ºC)
T
-200.0~400.0ºC
0.1ºC
N
-200.0~1300.0ºC
0.1ºC
Thermoresistance
Pt100
-200.0~850.0ºC
0.1ºC
±0.05% Setting value ±0.6ºC
Adopt the PT100-385 1mA
(Note 1.2)
Cu50
-50.0~150.0ºC
Note:
1. Not including the accessorial lead resistance.
2. The range of exciting current:0.5mA~2mA,the max.output voltage≤2V.
3. Not including the accuracy of inner temperature compensatory transducer.
The range of inner temperature compensatory transducer:-10-50°C,compensatory error ≤0.5°C.
4. Temperature coefficient: ±0.005%, range /°C(0°C-18°C,28°C-50°C).