No comment |
>
Instrumentation >
Digital multimeter >
Infrared Thermometer 0.5% repetition rate FLUKE561 FLUKEPLCAGE: Infrared Thermometer 0.5% rate FLUKE561 FLUKE New
Description
Fluke 561 IR and contact thermometer collection of industrial, electrical, HVAC technician needs a variety of industries the temperature measurement function, two kinds of both infrared and contact temperature measurement methods, one can use at most pieces. It's fast, efficient, easy to use and can save you valuable time and effort. Meanwhile, you can also choose their own needs and flexible test or contact infrared temperature measurement. The infrared thermometer can be completed in an instant heat, motion, charged, difficult to access the object measurement can be used to check the motor, insulation, breakers, radiant heat pipe, the connection point corrosion, line fault, and, without the ladders Scan directly HVAC and air conditioning pipes and other inaccessible objects. Measuring pipe, you can either use the handy Velcro pipe probe, you can also make use of their existing standard K-type mini-connector thermocouple probe to measure the internal temperature or overheating.
Features:
Near infrared thermometer can not only measure, but also
long-distance measurement, do not use a ladder.
Contact temperature measurement function is compatible
with all standard mini connector K-type thermocouple, you
absolutely do not need to purchase a separate thermocouple
for its products, thereby protecting your investment.
Velcro pipe probe can be used to heat, cold and other measuring
contact and non-contact temperature measurement.
Single-point laser sighting survey
Temperature measurement range -40 centigrade to550 centigrade, the vast majority of industrial and civil applications can use.
Very easy to adjust the emitted energy (high, medium, low), so as to more accurately measure the pipe surface is very smooth, and other materials.
Light weight (340 g), easy to carry
Easy usage charges
High efficiency - for infrared measurement, you do not turn off the device.
MIN, MAX, DIF functions help you quickly find the problem
No matter the size of the object, can quickly and efficiently complete failure scanning. Specifications:
Temperature Range
-40 To 550 centigrade
Display Resolution
Reading 0.1 centigrade
D: S (Distance to spot size)
12:1
Easy emissivity selector can be adjusted through three settings:
Low (0.3), medium (0.7) and high (0.95)
Display Accuracy (Assumes ambient operating temperature of 23 centigrade to 25 centigrade)
Reading or +/- 1 centigrade (whichever is greater), 0 centigrade or less is +/- 1 centigrade +/- 0.1 centigrade / 1 centigrade
Response Time
500 ms (95% of reading)
Repeatability
+/- 0.5% of reading or +/- 1 centigrade, whichever is greater
Spectral Response
8ohm to 14ohm
Laser sighting
Single-point laser
Laser Close
Laser at ambient temperature exceeds 40 centigrade when closed
Laser power
2 (II) class action; Output <1mW, wavelength 630-670 nm
Relative humidity
10% to 90% RH, non-condensing temperature <30 centigrade
Power
2 AA batteries (alkaline or nickel-cadmium batteries)
Battery life
12 hours
Display Hold
7 seconds
Backlit display
Yes, dual temperature display LCD (current temperature and MAX / MIN / DIF / KTC temperature), low battery indication, F / C unit indicator, and Scan / Hold options
No comment |