Milliamp Process Clamp Meter FLUKEF773 FLUKE NEWPLCAGE: Milliamp Process Clamp Meter FLUKEF773 FLUKE NEWFEATURES:
Best in class 0.2 % accuracy0.01 mA resolution and sensitivityMeasure 4 to 20 mA signals without ”breaking the loop”Measure mA signals for PLC and control system analog I/ODual backlit display with both mA measurement and percent of 4 to 20 mA spanMeasurement Spotlight illuminates hard to see wires in dark enclosuresDetachable clamp with extension cable for measurements in tight locationsMeasure 10 to 50 mA signals in older control systems using the 99.9 mA rangeAutomatically change the 4 to 20 mA output for remote testingPower saving features, auto-off, backlight timeout extend battery lifeHold function captures and displays changing measurementsMeasure 4 to 20 mA signals with in-circuit measurementSimultaneous mA in-circuit measurement with 24V loop power for powering and testing transmittersSource 4 to 20 mA signals for testing control system I/O or I/PsmA output linear ramp or 25% step outputAutomatically change the 4 to 20 mA output for remote testingDC voltage measurement to verify 24V power supplies or voltage I/O signalsSource DC voltage to test input devices that accept a 1-5V or 0 to 10V signalsScaled mA output provides a continuous mA signal that corresponds to the 4 to 20 mA signal measured by the mA clampEnables milliamp signal logging to a separate logging device without breaking the loopOutput mA signal enables a logging DMM or other device to record the 4- to 20 mA signal without breaking the loopmA in/out: simultaneously measure the mA signal with the clamp and source a mA signalApply a mA input signal to a device and measure its’ 4-20 mA output on devices such as valves or mA isolatorsVoltage output linear ramp or 25% step outputAutomatically change the voltage output for remote testing
What’s the difference between the 77X models?
*
Fluke-771
Fluke-772
Fluke-773
*
Non-contact mA loop measurement
Integrated mA loop troubleshooting
Advanced process troubleshooting
mA measure, jaw
mA measure, in circuit
*
Loop power
*
mA source
*
DCV measure
*
*
DCV source
*
*
mA scaled output
*
*
mA in / out
*
*
Environmental Specifications
Diameter of measurable conductor
0.177” or 4.5 mm max
Operating temperature
-10 to 50 ºC
Storage Temperature
-25 to 60ºC
Operating Humidity
< 90% @ <30 ºC, < 75% @ 30 to 55ºC
Operating Altitude
0 to 2000 m
IP Rating
IP 40
Size
44 x 70 x 246 mm (2 7/8 x 5 ¾ x 11 5/8 in)
Weight
410 g, (14.4 oz)
Vibration
Random 2 g, 5 to 500 Hz
Shock
1 meter drop test (except the jaw)
EMI/RFI
Meets EN61326-1
Note: For current measurement w/ JAW, add 1mA to
specification for EMC field strengths of 1V/m up to 3V/m.Temperature Coefficient
0.01% ºC
(<18ºC or > 28ºC)Power, battery life
(4) AA 1.5 V Alkaline, IEC LR6, 12 hours in mA source into 500 ohms
Warranty
Three years for electronics
one year for cable and clamp assembly
Functional Specifications
mA measurement
(Measured by clamp) 771/772/773Resolution and Range: 0 to 20.99 mA
Accuracy: 0.2%+2 counts
Resolution and Range: 21.0 mA to 100.0 mA
Accuracy: 1% + 5 countsmA measurement
(Measured in series with test jacks) 772/773Resolution and Range: 0 to 24.00 mA
Accuracy: 0.2% + 2countsmA source (Maximum mA Drive:
24 mA into 1,000 ohms)Resolution and Range: 0 to 24.00 mA
Accuracy: 0.2% + 2countsmA simulate
(Maximum voltage 50 V dc)Resolution and Range: 0 to 24.00 mA
Accuracy: 0.2% + 2countsVoltage source (2 mA maximum drive current)
Resolution and Range: 0 to 10.00 V dc
Accuracy: 0.2% + 2countsVoltatge Mesaure
Resolution and Range: 0 to 30.00 V dc
Accuracy: 0.2% + 2counts