VICTOR 14+ Temperaure CalibratorFeatures: High Accuracy of 0.02% for measure and source. 5 digits display for measure, and 6 digits display for source. A dual, Liquid Crystal; Display White LED for backlight Measuring Function
Function
Reference
Range
Resolution
Accuracy
Remark
DCV
50mV
-5.000mV?55.000mV
1μV
0.02+0.02
Input Resistan approximately 100MΩ
500mV
-50.00mV?550.00mV
10μV
0.02+0.01
OMH
500Ω
0.00Ω?550.00Ω
0.01Ω
0.05+0.02
Test current for 500Ω:Approximately
1mA;Test current for 5KΩ:
Approximately 0.1mA;Open circuit
voltage:About 2.5V;Dose not include
lead resistance;5KΩ
0.0000 KΩ?5.5000KΩ
0.1Ω
0.05+0.02
TC
R
0??1767?
1?
0?500? : 1.8?
By using ITS-90 temperature
scale;
The accuracy does not include
the error of internal temperature
compensation caused by a
sensor;S
0??1767?
500?1767? : 1.5?
K
-100.0??1372.0?
0.1?
-100.0?0.0? : 1.2?
E
-50.0??850.0?
0.0?1372.0? : 0.8?
J
-60.0??1120.0?
-50.0??0.0? : 0.9?
T
-100.0??400.0?
0.0??85.0? : 1.5?
N
-200.0??1300.0?
-60.0?0.0? : 1.0?
B
600??1820?
1?
0.0?1120.0? : 0.7?
L
-60.0??900.0?
0.1?
-100.0?0.0? : 1.0?
U
-100.0??600.0?
0.1?
0.0?400.0? : 0.7?
RTD
CONTINUITYPT100
385-200.0??800.0?
0.1?
-200.0?0.0? : 1.5?
By using temperature scale ITS-90.
Does not include lead resistance. Assuming all three RTD leads are matched for 3-w input.RTD
CONTINUITYPT1000
385-200.0??630.0?
0.0?1300.0? : 0.9?
PT200
385-200.0??630.0?
600?800? : 2.2?